Maintenance of EmiCo lite

Maintenance of EmiCo lite

EmiCo lite is a robust, low-maintenance gas analyzer designed for continuous operation in wastewater treatment plants. For consistently reliable measurements, a structured maintenance program is essential.Wastewater treatment plant operators already have many tasks to perform, and maintenance should not be one of them. That's why EmiCo lite requires only about 20 minutes of maintenance per year to maintain peak performance.

Calibration

Precise calibration is crucial for the reliability of the measured values. The EmiCo lite uses a semi-automatic calibration procedure with certified gas bags, which enables traceable and user-friendly calibration directly on site.

Calibration is carried out using the gas bags supplied (CO₂, CH₄, N₂O) and a 6 mm PTFE hose:

  • Zero point calibration with nitrogen (N₂)
  • Gas calibration (CO₂, CH₄, N₂O)
  • Return to measurement mode

Replacing the particle filters

The EmiCo lite is equipped with three particle filters that protect internal components from contamination and maintain measurement accuracy.

Exchange procedure:

  • Ensure a tight fit to guarantee gas tightness.
  • Remove the old filter and loosen the screw connections.
  • Align the new filters according to the flow direction (arrow on the filter).
  • Connect the hoses and secure them with coupling screws.
  • Ensure a tight fit to guarantee gas tightness.

Overview of maintenance plan

Zero point gas

To enable precise measurements, EmiCo lite determines the 0-point of the components. The system performs an automatic zero-point calibration with pure nitrogen every week. A nitrogen bottle is connected to the system for this purpose.

Semi-annual calibration

To avoid drifts, the EmiCo system must be calibrated every 6 months. Calibration is carried out by connecting gas bags with predetermined N₂O, CH₄ and CO₂ concentrations to the system. These bags are sent by Variolytics in advance.

Maintenance

The devices are monitored remotely by Variolytics to determine whether additional maintenance is required.

Service & Support

Variolytics offers maintenance contracts, calibration equipment, replacement filters and technical support to ensure the long-term performance of your EmiCo lite system.

FAQ – Frequently Asked Questions About EmiCo lite Maintenance

About 20 minutes per year. Every six months, you calibrate the device using pre-filled gas bags that we send you in advance—this takes less than 15 minutes. The zero point is automatically adjusted daily using nitrogen, and Variolytics monitors the devices remotely to determine whether additional maintenance is needed. Nothing in the wastewater needs to be cleaned.

EmiCo lite uses a semi-automatic calibration process with certified gas bags—traceable and performed directly on-site. You connect the included bags (CO₂, CH₄, N₂O) using a 6-mm PTFE tube. The device then performs a zero-point adjustment with nitrogen, calibrates each gas, and returns to measurement mode. This should be done every six months to prevent drift.

Daily. Every day, EmiCo lite automatically performs a zero-point calibration using pure nitrogen from its zero-gas cylinder—this ensures that the baseline for each gas remains precise without requiring any on-site intervention. In addition, the device is calibrated every six months using gas bags with known concentrations of N₂O, CH₄, and CO₂; this keeps the measurements free from drift.

We handle most of it—so your team spends as little time as possible on it. Variolytics monitors the equipment remotely, visits once a year for maintenance and calibration as part of the service contract, and supplies calibration gas pouches and replacement filters. Your team only needs to perform a few simple tasks in between, with our support. We ensure that the system runs smoothly and delivers reliable results.

Yes, every EmiCo lite comes with a service contract. It includes ongoing support and an annual visit from Variolytics for maintenance and calibration. This keeps the time commitment for your team to a minimum, and we ensure that the system runs smoothly and delivers reliable data. The costs are included in your customized quote.

Above all, the three particulate filters. They protect the internal components from contamination and ensure accurate measurements. To replace them, loosen the screw connections, remove the old filter, insert the new one in the direction of flow (indicated by the arrow on the filter), and reconnect the hoses so they are gas-tight. Variolytics supplies replacement filters; we can see when maintenance is due via the remote monitoring system.

Call the Variolytics service hotline at +49 711 2525 9620 or email service@variolytics.com. We can assist you with maintenance, calibration equipment, replacement filters, and technical questions about your EmiCo lite. For questions regarding IT security or to report a vulnerability, please email security@variolytics.com to ensure your message reaches the right team.

Through the online portal, we spot an error immediately and notify you—often before you even notice it yourself. The error code indicates what needs to be done: Either your team makes a minor adjustment with our assistance, or our service team schedules a fix. In either case, you don’t have to diagnose the issue on your own.